Probe Testing Adapter Procedure (WI-phx42-115)

Click here for a printable version of this procedure.

Items Needed

Part Name LDAR Part #
phx42 4200
Probe Testing Adapter 4300
phxApp for Self-check n/a

Perform the following steps whenever a filter error isn’t resolved by following on-screen prompts.

  1. Remove the probe, attach the probe testing adapter with a clean probe tip filter directly to the phx42, and run a self-check.

  2. If there are no errors, then the probe was the cause of the issue. Report the probe issue to LDARtools Support.

  3. If there are errors, submit the issue to LDARtools and comment: “Self-check completed with Probe Testing Adapter.”