Items Needed
Part Name | LDAR Part # |
---|---|
phx42 | 4200 |
Probe Testing Adapter | 4300 |
phxApp for Self-check | n/a |
Perform the following steps whenever a filter error isn’t resolved by following on-screen prompts.
-
Remove the probe, attach the probe testing adapter with a clean probe tip filter directly to the phx42, and run a self-check.
-
If there are no errors, then the probe was the cause of the issue. Report the probe issue to LDARtools Support.
-
If there are errors, submit the issue to LDARtools and comment: “Self-check completed with Probe Testing Adapter.”