Attach the SpanBox flow meter (LDAR# 1645) to the calibration tee.
-
Verify the output of each SpanBox port is set to at least 0.5 L/min if using only a phx42, and 2.0 L/min if using a phx21 or a mixture of both.
-
Ensure the flow meter is in an upright position.
-
In Cal5.0, select Manual Box Control. Check the box for the input and output port you are testing first.
-
Block the open leg of the calibration tee and read the flowmeter to verify it is set to the correct value.
-
Repeat the process for each input and output combination until all ports have been checked for all available gases.
-
If the flow is incorrect, adjust the flow regulator.