If you need to change your SpanBox5 settings for extension probe use, please follow the steps below:
Edit Settings in Customer Portal
-
Log in to the Customer Portal.
-
On the Welcome screen, choose Unit Settings.
-
Locate the onsite SpanBox by its serial number (SB5-xxx, SBL510/530-xxx) and click the Edit button.
-
Select the Multi Products tab, and then the Extension Probe section dropdown.
-
Under the Probe Type dropdown, select Extension.
-
In the Extension Probe Length field, enter the length of the probe (from the phx42 Quick Connect to the tip filter).
-
Click Save.
Confirm Setting Update on SpanBox
Note: The SpanBox must be connected to the network.
-
On the onsite SpanBox, select Settings on the home screen menu.
-
The Settings response should return a “Success” prompt, confirming the SpanBox is now set for extension probe use.
-
Under Daily Cal on the Warm-up screen, refer to the Ext. Probe, and Flow Rate dropdown.
-
Calibrate/PreCal/Drift as required