Verify the green light is illuminated on the power adapter of SpanBox base.
-
Verify the power cord is plugged into the SpanBox base.
-
Disconnect and reconnect the power cord into the SpanBox base.
-
Verify the USB cable is plugged into the correct USB port on the SpanBox tablet.
-
This is mainly for Chuwi tablets because they have two USB-C ports (one for power, one for data). The SpanBox USB needs to be connected to the lowest USB-C port (assuming the serial number SBHU is right-side up).
-
-
Unplug the USB cable from the SpanBox base and tablet.
-
Reconnect the USB cable to the tablet and SpanBox base.
-
Tap Manual Box Control, if the manual box loads, then toggle on a phx port and listen for a click. If there is no clicking sound, continue to the next step. If the solenoids click, the SpanBox is ready for DailyCal.
-
Note: A “Connection Error” may still be present on the Home screen.
-
-
If an error appears when tapping Manual Box Control, continue to the next step.
-
Restart Cal5.0 by tapping Restart on the Menu.
-
After the restart completes, repeat step 7.
SpanBox5 & Cal5.0
Was this article helpful? If not, please submit feedback below.